Wiretrol II Instruction Manual

  1. Assembling piston to bore: insert piston into the color coded (wide printed band near one end) end of the glass bore. This end has been flame polished to prevent damage to the teflon piston.

  2. Insert the Wiretrol into the fluid to be dispensed. Hold the bore between the thumb and third finger and operate the plunger with the index finger. Draw up thumb and third finger and operate the plunger with the index finger. Draw up the sample until the bottom of the piston intersects the desired calibration line.

  3. Dispense the sample by depressing the piston until the handle touches the bore. If dispensing into a diluent, simply swirling the core and tip in the fluid will remowe all of the sample.

  4. Accuracy: ±1% T. D.

  5. The short plunger is used where cross contamination is a problem. The plunger is aligned with the upper calibration mark and the sample is then drawn up to the lower calibration mark, one-half of the stated volume. When the plunger is depressed the sample is delivered. since the plunger has not contaminated the sample it may be reused.

Should you have any questions or wish to reorder please feel free to use our toll free line: outside of Pennsylvania 800-523-7480 or call collect within Pennsylvania 610-353-0200.

Drummond Scientific Company
500 Parkway, Box 700
Broomall, PA 19008
Hours of Operation 8:30 - 4:30 EST

Click here for Drummond Dealers (US - International)
1-800-523-7480
610-353-0200
Fax: 610-353-6204
email: info@drummondsci.com